User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 5200A Sub 2 DATE: 24-Oct-97 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 15 NUMBER OF LINES: 120 CONFIGURATION: Datron 1281 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {* Millivolt Range Tests *} 1.002 JMP 2.001 1.003 EVAL 2.001 DISP Remove the sense connection from the test setup. 2.002 DISP Change the UUT as follows: 2.002 DISP VOLTAGE RANGE ...................... 100 mV 2.002 DISP VOLTAGE dials ...................... 100.0000 2.002 DISP FREQUENCY RANGE .................... 1 KHz 2.002 DISP FREQUENCY dials .................... 1.000 2.003 DISP Set the UUT MODE switch to OBPR. 2.004 IEEE [@1281]RDG?[D5000][T0][I] 2.005 MEM* 1000 2.006 ACC 200 100.0000mV 100P% 10P/ 2.007 MEME 2.008 MEMC mV 0.048U #! Test Tol 0.048, Sys Tol 0.012, TUR 4.000 (>= 4.00). 3.001 DISP Set the UUT MODE switch to STBY. 3.002 DISP Change the UUT as follows: 3.002 DISP VOLTAGE RANGE ...................... 10 mV 3.003 DISP Set the UUT MODE switch to OBPR. 3.004 IEEE [@1281]RDG?[D5000][T0][I] 3.005 MEM* 1000 3.006 ACC 200 10.00000mV 100P% 10P/ 3.007 MEME 3.008 MEMC mV 0.012U #! Test Tol 0.012, Sys Tol 0.003, TUR 4.000 (>= 4.00). 4.001 DISP Set the UUT MODE switch to STBY. 4.002 DISP Change the UUT as follows: 4.002 DISP VOLTAGE dials ...................... 1.00000 4.003 DISP Set the UUT MODE switch to OBPR. 4.004 IEEE [@1281]RDG?[D5000][T0][I] 4.005 MEM* 1000 4.006 ACC 200 1.000000mV 100P% 10P/ 4.007 MEME 4.008 MEMC mV 0.0084U 10mVrange #! WARNING: Test Tol 0.0084, Sys Tol 0.0021, TUR 4.000 (< 4.00). 5.001 DISP Set the UUT MODE switch to STBY. 5.002 HEAD {* Voltage Linearity Test *} 5.003 JMP 6.001 5.004 EVAL 6.001 DISP Connect the UUT SENSE HI/LO to the Datron HI/LO inputs. 6.002 DISP Change the UUT as follows: 6.002 DISP VOLTAGE RANGE ...................... 10 V 6.002 DISP VOLTAGE dials ...................... 9.99999 6.002 DISP SENSE .............................. EXT 6.003 ACC 20 9.999990 60P% 10P/ 6.004 IEEE [@1281]RDG?[D5000][T0][I] 6.005 MEME 6.006 MEMC V 0.0032U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.0008, TUR 0.400 (< 4.00). 7.001 DISP Change the UUT as follows: 7.001 DISP VOLTAGE dials ...................... 8.88888 7.002 ACC 20 8.888880 60P% 10P/ 7.003 IEEE [@1281]RDG?[D5000][T0][I] 7.004 MEME 7.005 MEMC V 0.00293U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.00073333, TUR 0.436 (< 4.00). 8.001 DISP Change the UUT as follows: 8.001 DISP VOLTAGE dials ...................... 7.77777 8.002 ACC 20 7.777770 60P% 10P/ 8.003 IEEE [@1281]RDG?[D5000][T0][I] 8.004 MEME 8.005 MEMC V 0.00267U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.00066667, TUR 0.480 (< 4.00). 9.001 DISP Change the UUT as follows: 9.001 DISP VOLTAGE dials ...................... 6.66666 9.002 ACC 20 6.666660 60P% 10P/ 9.003 IEEE [@1281]RDG?[D5000][T0][I] 9.004 MEME 9.005 MEMC V 0.0024U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.0006, TUR 0.533 (< 4.00). 10.001 DISP Change the UUT as follows: 10.001 DISP VOLTAGE dials ...................... 5.55555 10.002 ACC 20 5.555550 60P% 10P/ 10.003 IEEE [@1281]RDG?[D5000][T0][I] 10.004 MEME 10.005 MEMC V 0.00213U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.00053333, TUR 0.600 (< 4.00). 11.001 DISP Change the UUT as follows: 11.001 DISP VOLTAGE dials ...................... 4.44444 11.002 ACC 20 4.444440 60P% 10P/ 11.003 IEEE [@1281]RDG?[D5000][T0][I] 11.004 MEME 11.005 MEMC V 0.00187U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.00046667, TUR 0.686 (< 4.00). 12.001 DISP Change the UUT as follows: 12.001 DISP VOLTAGE dials ...................... 3.33333 12.002 ACC 20 3.333330 60P% 10P/ 12.003 IEEE [@1281]RDG?[D5000][T0][I] 12.004 MEME 12.005 MEMC V 0.0016U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.0004, TUR 0.800 (< 4.00). 13.001 DISP Change the UUT as follows: 13.001 DISP VOLTAGE dials ...................... 2.22222 13.002 ACC 20 2.222220 60P% 10P/ 13.003 IEEE [@1281]RDG?[D5000][T0][I] 13.004 MEME 13.005 MEMC V 0.00133U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 0.00033333, TUR 0.960 (< 4.00). 14.001 DISP Change the UUT as follows: 14.001 DISP VOLTAGE dials ...................... 1.11111 14.002 ACC 2 1.111110 60P% 10P/ 14.003 IEEE [@1281]RDG?[D5000][T0][I] 14.004 MEME 14.005 MEMC V 0.00035U 1kH #! WARNING: Test Tol 0.00032, Sys Tol 8.6667e-005, TUR 3.692 (< 4.00). 15.001 DISP Set the UUT MODE switch to STBY. 15.002 DISP Remove all connections.